P.V. Method types
DoseControl has two different test methods for performance verification. The P.V. Method defines how your P.V. test will function. Each P.V. Method is configured with the reference standard information.
You choose which test method type works best for you.
P.V. Method: P.V. Method has one (1) testing wavelength for a photometric accuracy. Choose this test method to test a single wavelength near the wavelength of use. Example - if your target wavelength of use if 552nm for B3 film, you should use 546.1nm as the test wavelength.
• See how to Setup a basic P.V. Method.
• See how to Execute a P.V. test (basic method).
P.V. Method Bracket Wavelength of Use: Method with two (2) testing wavelengths for photometric accuracy. Choose this test method to bracket a measurement wavelength of use. Example – if your target wavelength of use is 552nm for B3 film, you should bracket this with 465nm and 590nm as the test wavelengths.
• See how to Setup a P.V. Method (Bracket Wavelength of Use).
• See how to Execute a P.V. test (bracket wavelength of use method).
View and Control P.V. Methods
The Performance Verification Tests screen displays the list of P.V. Methods setup in the software.
1 P.V. Method ID and Description
2 Default: Select one of the P.V. test methods as the default. The default option will appear at the top of the test options available to the user.
3 Is Active: Activate one or more P.V. test methods. The active P.V. test methods option will appear in the options available to the user.

Performance Verification Tests screen displays the P.V. Methods configured in the software.

The dropdown options that the user sees at the start of a new P.V. test.